Medical device recalls Moderate risk

Cook Inc. recalls Percutaneous Malecot Nephrostomy Set, Global Product No. G16714 G19107 G19108

Recall date
January 4, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0968-2020
FDA classification
Class II
Brand / firm
Cook Inc.
Sold / distributed
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia,…

Why it was recalled

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Percutaneous Malecot Nephrostomy Set, Global Product No. G16714 G19107 G19108