Medical device recalls Moderate risk

Cook Inc. recalls Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22…

Recall date
August 31, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0105-2021
FDA classification
Class II
Brand / firm
Cook Inc.
Sold / distributed
Worldwide distribution - US Nationwide.

Why it was recalled

The indwell time for the Percutaneous Neonatal Pigtail Nephrostomy Set and Pediatric Nephrostomy Stent Set is decreasing from 4 months to 4 weeks. The current IFU is in process of being updated and will be provided with orders placed following implementation.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Percutaneous Neonatal Pigtail Nephrostomy Set, Reference Part Number 080106, Order Number G14508. The set contains a 22 gage, 5 cm Chiba Needle; an 18 gage, 5 cm trocar needle; an .038-inch (.97 mm) diameter, 30 cm PTFE-coated stainless steel; a 5, 6, and 7 French, 10 cm radiopaque dilator; a 6 French, 7 cm radiopaque pigtail catheter; a retention disc with pull tie; and a 10 French, 10 cm clear, polyvinylchloride connecting tube with one-way stopcock.

Get recall alerts

Free email alert whenever Cook Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Cook Inc.