Medical device recalls Moderate risk

Cook Inc. recalls Slip-Cath Beacon Tip Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic pr…

Recall date
April 15, 2016
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2614-2016
FDA classification
Class II
Brand / firm
Cook Inc.
Sold / distributed
Worldwide Distribution - US (Nationwide) and Internationally to Algeria, Andorra, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Barbados, Belgium, Bermuda, Brazil, Brunei Darussalam, Bulgaria, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Cyprus, Czech Republic, Denmark…

Why it was recalled

Increase in reports of polymer degradation of the catheter tip, resulting in tip fracture and/or separation. degradation could include loss of device function, separation of a device segment leading to medical intervention, or complications resulting from a separated segment. Such complications include device fragments in the vascular system, genitourinary system, or other soft tissues. Fragments within the vascular system could result in embolization to the heart or lungs, or occluding blood flow to end organs.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Slip-Cath Beacon Tip Catheter Catheter, Intravascular, Diagnostic The catheters are intended for use in angiographic procedures by physicians trained and experienced in angiographic techniques.

Get recall alerts

Free email alert whenever Cook Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Cook Inc.