Medical device recalls Moderate risk

Copan Diagnostics Inc recalls COPAN REF: 2U045S01 300uL Transfer Pipette Individually wrapped pipettes, 1000 pcs.

Recall date
July 14, 2023
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2620-2023
FDA classification
Class II
Brand / firm
Copan Diagnostics Inc
Sold / distributed
US: CA OUS: None

Why it was recalled

Due to plastic defects, the pipette aspiration capability potentially doesn't meet the expected tolerance

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

COPAN REF: 2U045S01 300uL Transfer Pipette Individually wrapped pipettes, 1000 pcs.

Get recall alerts

Free email alert whenever Copan Diagnostics Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Copan Diagnostics Inc