Copan WASP recalls PhenoMATRIX
- Recall date
- May 6, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1812-2025
- FDA classification
- Class II
- Brand / firm
- Copan WASP
- Sold / distributed
- US: Nationwide distribution in the states of UT, AR, CA, PA, MD, NC, IL, MA, NJ, KS, TX, NY, NV.
Why it was recalled
AI-powered bacterial culture plate interpretation and workup software does not have Premarket Approval or pre-market notification (510(k)/de novo), so there is potential risk that the colony count and/or morphology information from images may not accurately reflect actual plate and may lead to inaccurate counts and/or morphology characteristics impacting patient diagnosis and subsequent care.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
PhenoMATRIX