Medical device recalls Moderate risk

Corin Ltd recalls Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity…

Recall date
January 25, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1291-2021
FDA classification
Class II
Brand / firm
Corin Ltd
Sold / distributed
Worldwide distribution - US Nationwide distribution and the countries of Colombia, Poland, Germany, Australia.

Why it was recalled

The size indicated on the labeling on the outer packaging, on the inner packaging, and on the patient stickers inside the packaging is incorrect: the Trinity Biolox Delta Ceramix Head 32XL is labeled as 36XL and vice-versa.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Corin BIOLOX DELTA MOD HEAD 36mm EX LONG +8mm12/14 TAPER. Corin Trinity Acetabular System with Extra Long Heads/Trinity Biolox Delta Ceramic Head (Generic Name: Hip Prosthesis). The Trinity Acetabular System is a modular acetabular cup system consisting of a press fit, titanium alloy shell; acetabular liners in neutral offset, +4mm offset, +4mm oblique, neutral 4mm EPW; and ceramic and CoCrMo modular heads. The ceramic and CoCrMo modular heads are compatible with Corin titanium femoral stems. The Trinity Biolox Ceramic heads are packaged in a double peel pouch configuration.

Get recall alerts

Free email alert whenever Corin Ltd has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Corin Ltd