Medical device recalls Moderate risk

Covidien Llc recalls Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-c…

Recall date
April 30, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1993-2020
FDA classification
Class II
Brand / firm
Covidien Llc
Sold / distributed
US Nationwide distribution.

Why it was recalled

Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dover 400 ml 100% Silicone Silver-Hydrogel Coated Premium Urine Meter Foley Tray,18 Fr/Ch (6.0 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Temperature Sensor, Sterile, Product Code: P4P18XTS - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.