Medical device recalls Moderate risk

Covidien LLC recalls HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03…

Recall date
January 30, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1557-2026
FDA classification
Class II
Brand / firm
Covidien LLC
Sold / distributed
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, FL, GA, IL, IN, KY, LA, MI, MN, MS, NC, NY, OH, OK, OR, SC, TN, TX, and VA. The country of Spain.

Why it was recalled

Due to complaints and investigations stating that alarms from the primary patient bedside monitor was not alarming/being transmitted and received on the remote patient monitoring system.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

HealthCast "Vital Sync" Remote Patient Monitoring System which consists of: Product Number: VSLBASE03P / VITALSYNCSW03 Model / UDI-DI codes: Vital Sync 3.3.1 / 10884521844933 Vital Sync 3.4.0 / 10884521844940 Vital Sync 3.4.1 / 10884521847569 (OUS Only) Vital Sync 3.4.2 / 10884521856127 Product Number: VSLBASE04P / VITALSYNCSW04 Model / UDI-DI codes: Vital Sync 4.0.0 / 10884521849358 Vital Sync 4.0.2 / 10884521856134 The Vital Sync" Informatics Manager is software that is intended to route and store medical device data and device diagnostic information from supported devices to the Virtual Patient Monitoring Platform, 3rd Party Annunciation Systems, Electronic Medical Record (EMR) and Clinical Information System (CIS).

Get recall alerts

Free email alert whenever Covidien LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Covidien LLC