Covidien, LP recalls Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
- Recall date
- September 25, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0485-2026
- FDA classification
- Class II
- Brand / firm
- Covidien, LP
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Chile, Estonia, Finland, France, Germany, Greece, Hungary, Israel, Italy, Japan, Kuwait, Malaysia, Netherlands, Poland, Portugal, Romania, Saudi Arabia, Slovakia, Spain, United Kingdom.
Why it was recalled
A manufacturing issue could lead to a malformed cover tube press which can lead to internal components' movement during firing - leading to the potential for uncontrolled motion.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Covidien Signia" Small Diameter Curved Tip Intelligent Reload (45mm Vascular/Thin); CFN: SIGSDL45CTVT;
Get recall alerts
Free email alert whenever Covidien, LP has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Covidien, LP