CTL Medical Corporation recalls CTL Medical Hex Driver MIS, 1/4" Square QC, 4.0 Ratcheting Ref 019.7050
- Recall date
- July 5, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1031-2019
- FDA classification
- Class II
- Brand / firm
- CTL Medical Corporation
- Sold / distributed
- distributor and user level
Why it was recalled
the anti-toggle ring of the hex driver may separate from the driver shaft. If this occurs during surgery, the anti-toggle ring can be pressed into the pedicle screw head and cause separation of the screw from the housing upon tightening. if the anti-toggle ring remains in the pedicle crew head and can interfere with final tightening of the pedicle screw construct. this can lead to failure of the set crew and relies of the rod from the construct.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
CTL Medical Hex Driver MIS, 1/4" Square QC, 4.0 Ratcheting Ref 019.7050