Medical device recalls Moderate risk

Cyberonics, Inc recalls VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizu…

Recall date
December 18, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0659-2016
FDA classification
Class II
Brand / firm
Cyberonics, Inc
Sold / distributed
Worldwide Distribution -- United States, Austria, Belgium, Croatia, Czech Republic, Finland, France, Germany, Iceland, Italy, Netherlands, Norway, Poland, Slovakia, Spain, Sweden, Switzerland, United Kingdom, Cyprus, Israel, Kuwait, Lebanon, Oman, Qatar, Saudi Arabia, and United Arab Emirates.

Why it was recalled

Recall being initiated in response to three reports of "Burst Watchdog Timeout" events occurring with the Model 106 AspireSR Generator, resulting in a device reset condition where stimulation output is disabled.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VNS Therapy AspireSR Generator Model 106. Indicated for use as an adjunctive therapy in reducing the frequency of seizures.