Datascope Corp. recalls CS100 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3013-55 For cardiovascular use.
- Recall date
- June 17, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1942-2019
- FDA classification
- Class I
- Brand / firm
- Datascope Corp.
- Sold / distributed
- Worldwide distribution. US Nationwide, AFGHANISTAN , HUNGARY , PANAMA , ALGERIA , ICELAND , PARAGUAY , ARGENTINA , INDIA , PERU , AUSTRALIA , INDONESIA , PHILIPPINES , AUSTRIA , IRAN , POLAND , AZERBAIJAN , IRAQ , PORTUGAL , BAHRAIN , IRELAND , QATAR , BANGLADESH , ISRAEL , ROMANIA , BELARUS , ITAL…
Why it was recalled
If battery maintenance is not performed appropriately, the battery may provide less than the minimum run time of operating power. Revised labeling and training are being provided.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
CS100 Intra-Aortic Balloon Pump, English, 220V; 0998-00-3013-55 For cardiovascular use.