Datascope Corporation recalls MEGA¿ 7.5Fr. 30cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ MEGA¿ 7.5Fr. 40cc Intra-Aortic Balloon C…
- Recall date
- May 13, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1935-2016
- FDA classification
- Class II
- Brand / firm
- Datascope Corporation
- Sold / distributed
- Worldwide Distribution.
Why it was recalled
It was discovered during post-quarantine packaging on 2/16/2016 that the IFU for the STATLOCK Sheath Stabilization device for Percutaneous Sheath Introducer (Vendor PN 2403097-1405R) was erroneously packaged with the Mega and Sensation Plus IAB Catheter Kits instead of the IFU for the STATLOCK Catheter Stabilization device for IAB catheters (Vendor PN 24030905-1406R).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
MEGA¿ 7.5Fr. 30cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ MEGA¿ 7.5Fr. 40cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ MEGA¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ MEGA¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit, STATLOCK¿ & APA
Get recall alerts
Free email alert whenever Datascope Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Datascope Corporation