Medical device recalls Moderate risk

Datascope Corporation recalls SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-0…

Recall date
April 12, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2164-2017
FDA classification
Class II
Brand / firm
Datascope Corporation
Sold / distributed
Nationwide

Why it was recalled

Maquet/Getinge is initiating a voluntary product removal involving four lot numbers of SENSATION PLUS 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK since during internal pressure (leak) testing of the SENSATION PLUS 8Fr. 50cc, IABC, failures related to the tip seal leak were identified.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

SENSATION PLUS¿ 8Fr. 50cc Intra-Aortic Balloon Catheter with Insertion Kit & STATLOCK¿ Product Code/Part Number: 0684-00-0576-01U is a cardiac assist device which support the heart's left ventricle by increasing coronary perfusion and reducing the work of the left ventricle.

Get recall alerts

Free email alert whenever Datascope Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Datascope Corporation