Degania Silicone, Ltd. recalls Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN JP-2190 An introduction/drainage cath…
- Recall date
- June 28, 2019
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2263-2019
- FDA classification
- Class II
- Brand / firm
- Degania Silicone, Ltd.
- Sold / distributed
- Nationwide Distribution
Why it was recalled
15FR and 19FR round fluted drain channel drain may break during use (mostly during removal) in connection point between clear tube and white profile part. requiring additional surgical intervention to remove the broken end from patient's body
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Cardinal Health Jackson-Pratt Channel Drain 19FR, Full flutes Cardinal Health: PN JP-2190 An introduction/drainage catheter is a device that is a flexible single or multilumen tube intended to be used to introduce nondrug fluids into body cavities other than blood vessels, drain fluids from body cavities, or evaluate certain physiologic conditions.
Get recall alerts
Free email alert whenever Degania Silicone, Ltd. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Degania Silicone, Ltd.