Medical device recalls Moderate risk

DePuy Orthopaedics, Inc. recalls ATTUNE CONV RP PS TB TRL SZ10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim componen…

Recall date
June 12, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2144-2015
FDA classification
Class II
Brand / firm
DePuy Orthopaedics, Inc.
Sold / distributed
Worldwide Distribution - US including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY and Internationally to CANADA , AUSTRALIA, AUSTRIA, BELGIU…

Why it was recalled

Potential for the Balseal, a small wire spring coil located on the post feature of the ATTUNE Knee Tibial Articulation Surface Instruments used in trialing to become damaged and disassociate (come off). If the Balseal is separated from the post, it has the potential to enter the surgical site and be left in the patient if the surgeon is unaware of the disassociation.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

ATTUNE CONV RP PS TB TRL SZ10 INTENDED USE: The Tibial Articulation Surface Trial snaps together with the Shim component to function as the Insert Trial during a total knee replacement. A Balseal is attached to each of the two post features on the Articulation Surface to provide a connection force between the Tibial Articulation Surface and Shim to ensure secure engagement between the components.