DePuy Orthopaedics, Inc. recalls GLOBAL UNITE STD STEM SZ 10 Part Number: 110010100

Recall date
December 7, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0486-2022
FDA classification
Class II
Brand / firm
DePuy Orthopaedics, Inc.
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Austria, Australia, Belgium, Canada, Denmark, Estonia, Finland, France, Germany, Greece, Ireland, Israel, Italy, Netherlands, New Zealand, Norway, Poland, South Africa, Sweden, Switzerland, Thailand, United Kingdom.

Why it was recalled

Stems may have an undersized spigot bore diameter therefore separating the proximal body from the stem may not be possible without surgical delay and/or bone damage

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GLOBAL UNITE STD STEM SZ 10 Part Number: 110010100