Medical device recalls Moderate risk

DePuy Orthopaedics, Inc. recalls The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replac…

Recall date
November 28, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1012-2018
FDA classification
Class II
Brand / firm
DePuy Orthopaedics, Inc.
Sold / distributed
Distribution Level: The device removal is extended to the companies US Distributors/Sales Offices and Medical Facilities that may have used, purchased, or received the affected devices. Approximately 53 units distributed to firm's international affiliates.

Why it was recalled

The screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LE¿O-1100-30-000 REV D.pdf.pdf, LE¿O-1100-40-000 REV D.pdf.pdf, LE¿O-110030100_REV D.pdf.pdf, LE¿O-110030110_REV D.pdf.pdf, LE¿O-110030120_REV D.pdf.pdf