Medical device recalls Moderate risk

DePuy Orthopaedics, Inc. recalls VELYS Robotic-Assisted Solution Base Product No.: 451570100

Recall date
March 11, 2022
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1029-2022
FDA classification
Class II
Brand / firm
DePuy Orthopaedics, Inc.
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Israel, Australia & New Zealand.

Why it was recalled

System software v1.5.1 has a system software issue related to the Daylight Savings Time (DST) change that can cause a system error, requiring the user to restart the system and potentially cause a delay in treatment

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VELYS Robotic-Assisted Solution Base Product No.: 451570100