Medical device recalls Moderate risk

Dexcom Inc recalls Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is…

Recall date
February 3, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1560-2020
FDA classification
Class II
Brand / firm
Dexcom Inc
Sold / distributed
World-wide Distribution

Why it was recalled

The company identified potential interference from hydroxyurea. Patient use of the anti-neoplastic drug may falsely elevate glucose readings on the Continuous Glucose Monitoring Systems.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dexcom G4 PLATINUM Continuous Glucose Monitoring System The Dexcom G4 PLATINUM Continuous Glucose Monitoring System is a glucose monitoring device indicated for detecting trends and tracking patterns in persons (age 18 and older) with diabetes.

Get recall alerts

Free email alert whenever Dexcom Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Dexcom Inc