Medical device recalls Moderate risk

Diagnostic Hybrids, Inc. recalls MicroVue CIC-C1q EIA, Model A001. An enzyme immunoassay for the detection of circulating immune complexes in human seru…

Recall date
April 4, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1496-2018
FDA classification
Class II
Brand / firm
Diagnostic Hybrids, Inc.
Sold / distributed
The products were distributed to the following US states: NY, TX, CO, UT. The products were distributed to the following foreign countries: Switzerland.

Why it was recalled

There is a potential for the slope of the standard curve to fail to meet the assay validation requirements for the assay.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MicroVue CIC-C1q EIA, Model A001. An enzyme immunoassay for the detection of circulating immune complexes in human serum or plasma. The assay uses highly pure, functional human C1q coated in the solid phase to capture immune complexes (CIC). In the first stage, CIC in the diluted patient samples and HAGG in the controls and standards are dispensed into the C1q coated assay wells. After incubation, unbound material is removed in a washing step and a ready to use conjugate (goat anti-human Ig-HRP) is added. After a second incubation, unbound conjugate is washed away. After addition of a substrate and a short incubation interval, the quantity of CIC in the sample (¿g Eq/ml) can be determined by comparison to a standard curve.

Get recall alerts

Free email alert whenever Diagnostic Hybrids, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Diagnostic Hybrids, Inc.