DIXI MEDICAL USA recalls The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0…
- Recall date
- December 9, 2021
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0536-2022
- FDA classification
- Class II
- Brand / firm
- DIXI MEDICAL USA
- Sold / distributed
- US Nationwide distribution.
Why it was recalled
Potential deformation of the internal packaging (blister pack) could compromise sterile barrier of the device when removed from the Tyvek bag
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
The DIXI Medical Microdeep Depth Electrode (Microdeep Depth Electrode or Electrode) is a smooth electrode that is 0.8 mm in diameter with a rounded (hemispherical) tip. The electrode has multiple platinum/iridium contacts that are 2 mm long, with a total exploration length varying from 16 mm to 80.5 mm according to the electrode reference. Microdeep Depth Electrodes are intended for temporary use with recording, monitoring and stimulation equipment for the recording, monitoring and stimulation of electrical signals at the subsurface level of the brain.