Medical device recalls Moderate risk

Draegar Medical Systems, Inc. recalls Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722

Recall date
June 6, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2734-2017
FDA classification
Class II
Brand / firm
Draegar Medical Systems, Inc.
Sold / distributed
AZ, NH, SD, TN, TX Foreign: Australia Austria Belgium Canada China Costa Rica France Germany Great Britian Hong Kong India Ireland Italy Japan Malaysia Namibia Netherlands Poland Portugal Russian Fed. South Africa South Korea Spain Sweden Switzerland Thailand Turkey United Kingdom

Why it was recalled

Cockpits with revision index 06 or higher that contain 4GB RAM modules may not annunciate audio or visual alarms on the Cockpit and Central Station.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Draeger Medical Systems Infinity Acute Care System (M540) Catalog Numbers: MK31501/MK31701/MK31722