Medical device recalls Moderate risk

Draeger Medical, Inc. recalls Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous flow anesthesia system. The Apollo…

Recall date
May 24, 2018
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2253-2018
FDA classification
Class II
Brand / firm
Draeger Medical, Inc.
Sold / distributed
US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.

Why it was recalled

The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Apollo Anesthesia Machine; Cat. no. 8605310 Product Usage: Indicated as a continuous flow anesthesia system. The Apollo may be used for manually assisted or automatic ventilation, delivery of gases and anesthetic vapor, and monitoring of oxygen and CO2 concentration, breathing pressure, respiratory volume, and anesthetic agent concentration and identification. It is an inhalation anesthesia machine for use in operating, induction, and recovery rooms. It can be used with rebreathing systems, semiclosed to virtually closed systems with low flow and minimal flow techniques, and non-rebreathing systems (with the Auxiliary Common Gas Outlet). It may be used with O2, N2O, and Air supplied by a medical gas pipeline system or by externally mounted gas cylinders. Anesthetic agent can be delivered via vaporizers mounted on the machine. The Apollo is equipped with a compact breathing system, providing fresh-gas decoupling, PEEP, and pressure limitation. It has an electrically driven and electronically controlled ventilator.