Medical device recalls High risk

Draeger Medical, Inc. recalls Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation inten…

Recall date
October 15, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0668-2025
FDA classification
Class I
Brand / firm
Draeger Medical, Inc.
Sold / distributed
No Atlan A300 or A300XL devices were imported into the US. Devices were direct shipped from Draegerwerk Ag and Co. KGaA to the following countries: Argentina, Australia, Austria, Azerbaijan, Bahrain, Belgium, Brazil, Bulgaria, Cambodia, Canada, Chili, China, Colombia, Croatia, Czech Republic, Egypt…

Why it was recalled

The Atlan anesthesia workstation either indicated a failure of the piston ventilator before use or suffered a failure of the mechanical ventilation during use.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Brand Name: Atlan Product Name: Atlan A300XL Model/Catalog Number: 8621400 The Atlan is an anesthesia workstation intended for use in anesthetizing adults, pediatrics, and neonates.