Draeger Medical, Inc. recalls Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
- Recall date
- February 4, 2016
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0827-2016
- FDA classification
- Class I
- Brand / firm
- Draeger Medical, Inc.
- Sold / distributed
- US Nationwide Distribution including Puerto Rico.
Why it was recalled
The battery capacity of the optional PS500 power supply unit for the Infinity (ACS) Workstation Critical and Neonatal Care was rapidly and unexpectedly reduced, despite using the most recent power supply firmware version 1.50 released in November 2015 (recall Z-0436-2016). The batteries were prematurely discharged, even though an adequate battery charge status was displayed on the device.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
Evita Infinity V500 Ventilator; Babylog Infinity VN500 Ventilator. Ventilation.
Get recall alerts
Free email alert whenever Draeger Medical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Draeger Medical, Inc.