Draeger Medical, Inc. recalls Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults, ch…
- Recall date
- May 24, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2254-2018
- FDA classification
- Class II
- Brand / firm
- Draeger Medical, Inc.
- Sold / distributed
- US Nationwide Distribution - Healthcare facilities in AK, AL, AZ, CA, CO, FL, GA, IA, IL, IN, MA, MD, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, WA, WI, and WV.
Why it was recalled
The Draeger anesthesia device may be able to dose 100% N2O. In the event of a fault, the S-ORC module would not prevent setting an N2O flow that would result in a hypoxic mixture from being dosed to the patient. Potential adverse outcomes include death of the patient.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Perseus A500 Anesthesia Machine; Cat. no. MK06000 Product V Product Usage: Intended for use in anesthetizing adults, children, and neonates and can be used for automatic and manual ventilation, pressuresupported spontaneous breathing, and spontaneous breathing. Perseus is equipped with airway monitoring, gas measurement and device monitoring, O2 insufflation, and an anesthetic gas receiving system. Anesthesia is achieved through a mixture of pure oxygen and Air (medical compressed air) or pure oxygen and nitrous oxide, with the addition of volatile anesthetic agents. Ventilation is accomplished on the patient through a laryngeal mask airway, a full-face mask, or an endotracheal tube. The integrated breathing system can be used with partial rebreathing (low-flow or minimum-flow). A non-rebreathing system such as the Kuhn or Medec Water System may be used at the external fresh-gas outlet (optional). Perseus is specified for inhalational anesthesia and/or patient ventilation in accordance with the intended use during surgical or diagnostic interventions.
Get recall alerts
Free email alert whenever Draeger Medical, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Draeger Medical, Inc.