Medical device recalls Moderate risk

Dynavox Systems Llc recalls Dynavox T10 Speech Generating Device ( including Batteries)

Recall date
December 14, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1304-2018
FDA classification
Class II
Brand / firm
Dynavox Systems Llc
Sold / distributed
Nationally

Why it was recalled

The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Dynavox T10 Speech Generating Device ( including Batteries)