Dynavox Systems Llc recalls Dynavox T15 Speech Generating Device ( including Batteries)
- Recall date
- December 14, 2017
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1305-2018
- FDA classification
- Class II
- Brand / firm
- Dynavox Systems Llc
- Sold / distributed
- Nationally
Why it was recalled
The battery is integral to the device. If the device sits unused and is not charged for a lengthy period of time (a few months) the battery will go into a deep discharge state and the battery will expire. The battery may out-gas which may be described as swelled or puffed, as gas is trapped inside the pack surrounding the cells. This battery may cause the device case to separate a small amount.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Dynavox T15 Speech Generating Device ( including Batteries)