Medical device recalls Moderate risk

Electro Medical Systems SA recalls Brand Name: CLEANER BOTTLE Product Name: CLEANER BOTTLE Model/Catalog Number: EG-1000 Software Version: N/A Product Des…

Recall date
March 13, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1686-2024
FDA classification
Class II
Brand / firm
Electro Medical Systems SA
Sold / distributed
Worldwide distribution - U.S. Nationwide in the states of IL and TX. The countries of ALBANIA, ARMENIA, AUSTRALIA, AZERBAIJAN, BAHRAIN, BELGIUM, BRAZIL, BULGARIA, CANADA, CHINA, CZECH REPUBLIC, EGYPT, FRANCE, GEORGIA, GERMANY, GREECE, HONG KONG, HUNGARY, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZ…

Why it was recalled

Received complaints that bottles assembled with one batch of bottle collars can be disassembled in two parts when used under pressure (during treatment or storage if the use is not releasing the pressure inside the bottles).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Brand Name: CLEANER BOTTLE Product Name: CLEANER BOTTLE Model/Catalog Number: EG-1000 Software Version: N/A Product Description: CLEANER bottles used with EMS AIRFLOW PROPHYLAXIS MASTER / EMS AIRFLOW ONE devices to clean the water lines. The bottle is placed on the device during cleaning. Component: YES. AIRFLOW PROPHYLAXIS MASTER and AIRFLOW PROPHYLAXIS ONE

Get recall alerts

Free email alert whenever Electro Medical Systems SA has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Electro Medical Systems SA