Medical device recalls Moderate risk

Elekta Inc recalls Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.

Recall date
September 1, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0234-2018
FDA classification
Class II
Brand / firm
Elekta Inc
Sold / distributed
Worldwide Distribution - US Nationwide in the states of: VA, CA, TX, MO, NY, WA, NM, MA, FL, CT, OK, SC, MS, NY, GA, NC and the countries of: AUSTRALIA, BANGLADESH, BRAZIL, BULGARIA, CANADA, COLOMBIA, CZECH REPUBLIC, ECUADOR, FRANCE, GERMANY, GREECE, HONG KONG, INDIA, IRELAND, ITALY, JAPAN, JORDAN,…

Why it was recalled

Incorrect DICOM mapping of the exported collimator or couch angles from ERGO, which would lead to incorrect rotation of the collimator or couch when using a MLC device for planning.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Elekta ERGO++ Product Usage: ERGO is often used for stereotactic treatments.

Get recall alerts

Free email alert whenever Elekta Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Elekta Inc