Elekta, Inc. recalls Elekta Precise Treatment Table
- Recall date
- January 29, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1238-2015
- FDA classification
- Class II
- Brand / firm
- Elekta, Inc.
- Sold / distributed
- Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, E…
Why it was recalled
It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical mistreatment may occur if a position error is not detected.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Elekta Precise Treatment Table
Get recall alerts
Free email alert whenever Elekta, Inc. has a new recall — straight from official government data. Unsubscribe anytime.