Elekta, Inc. recalls Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as d…
- Recall date
- January 7, 2015
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1088-2015
- FDA classification
- Class II
- Brand / firm
- Elekta, Inc.
- Sold / distributed
- Worldwide Distribution - USA including AZ, CA, FL, KS, LA, MN, MO, NY, NC and Internationally to Austria, Bulgaria, Canada, China, Germany, Hungary, Italy, Malaysia, Malta, Netherlands, Norway, Russia, Spain, Sweden, Taiwan and Ukraine.
Why it was recalled
Elekta has identified a batch of securing bolts that fix the Gantry counter frame weight stack may fail.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Precise Digital Accelerator Intended to be used for radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.