Medical device recalls Moderate risk

Encore Medical, Lp recalls EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096

Recall date
February 4, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0611-2020
FDA classification
Class II
Brand / firm
Encore Medical, Lp
Sold / distributed
CO, WA, CA, UT, CA, TN, TX, MS, MO, AR, OK, LA, MN, IL, IN, KY, OH, MI, PA, SC, FL, PR, AL, GA, NM, AZ, NV, VA, RI, MD, ME PA, NY OUS: None

Why it was recalled

The firm has received reports/complaints concerning the breakage of the tips of the locking femoral Impactor. Additionally, if the impactor is not seated correctly on the implant prior to impaction, this may cause off axis loading which can increase the risk of device failure. This could cause a delay in surgery. If the tip is not seen during surgery, long term risks may include an inflammatory response, infection, and/ or pain

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

EMPOWR Locking Femoral Impactor , Part # 800-05-035, packaged in instrument tray FA EMP PAT TOOL KIT Part # 800-99-096