Medical device recalls Moderate risk

Encore Medical, LP recalls EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.

Recall date
June 19, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0479-2021
FDA classification
Class II
Brand / firm
Encore Medical, LP
Sold / distributed
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.

Why it was recalled

There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.

Get recall alerts

Free email alert whenever Encore Medical, LP has a new recall — straight from official government data. Unsubscribe anytime.