Medical device recalls Moderate risk

Encore Medical, Lp recalls EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.

Recall date
December 12, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0327-2018
FDA classification
Class II
Brand / firm
Encore Medical, Lp
Sold / distributed
One medical device distributor in California.

Why it was recalled

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.