Encore Medical, Lp recalls Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.
- Recall date
- September 19, 2018
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2090-2019
- FDA classification
- Class II
- Brand / firm
- Encore Medical, Lp
- Sold / distributed
- US ( MS, NY, MO, OK, UT, AZ)
Why it was recalled
Manufacturing error, the drill bit shaft is unable to properly capture the drill bit pin. Failure of the drill bit shaft to capture the drill bit pin could result in the drill bit coming loose which could result in a delay in surgery or adverse patient reaction resulting in a revision surgery.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Flexible Drill Shaft, FMP Acetabular System Instrumentation REF 803-05-239 RX Encore Medical, L.P.