Medical device recalls Moderate risk

Encore Medical, Lp recalls MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S…

Recall date
August 4, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-3102-2017
FDA classification
Class II
Brand / firm
Encore Medical, Lp
Sold / distributed
OH, VA

Why it was recalled

It was reported that the models used to manufacture S-200956/958 had the wrong dimension used to create the spherical radius cutting surface. Device design allows a larger spherical radius cutting surface than intended.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MODIFIED/SPECIAL INSTRUMENT, TURON GLENOID REAMER, REAM AND RUN, Catalog Numbers: REF S-200956 - 5mm PIN, SIZE 46 REF S-200957 - 5mm PIN, SIZE 50 REF S-200958 - 5mm PIN, SIZE 58

Get recall alerts

Free email alert whenever Encore Medical, Lp has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Encore Medical, Lp