Medical device recalls High risk

Endologix, Inc. recalls Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB26…

Recall date
May 6, 2020
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2263-2020
FDA classification
Class I
Brand / firm
Endologix, Inc.
Sold / distributed
Worldwide DIstribution: U.S (nationwide).: PA, AZ, NY, FL, HI, CA, OR, IL, MS, AR, WI, KY, TX, TN, MO, NJ, MA, MT, VA, MI, CO, WV, NC, NH, NV, CT, ME, WY, OH, OK, IN, UT, AL, WA, GA, SC, KS, LA, NM, MD, DC, NE, SD, RI, IA, VT, ND. and Countries of:: Argentina, Australia, Chile, Hong Kong, Japan, Ma…

Why it was recalled

A material weakness adjacent to the polymer fill channel may become compromised during pressurization with liquid polymer. Polymer leaks may occur during implantation of abdominal stent graft system. Clinical events related to polymer leaks may be systemic and/or aneurysm related.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Ovation iX Abdominal Stent Graft Systems, REF: TV-AB2080-I, TV-AB2080-J, TV-AB2380-I, TV-AB2380-J, TV-AB2680-I, TV-AB2680-J, TV-AB2980-I, TV-AB2980-J, TV-AB3480-I, TV-AB3480-J