Medical device recalls Moderate risk

Enztec Limited recalls REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE

Recall date
August 16, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0612-2020
FDA classification
Class II
Brand / firm
Enztec Limited
Sold / distributed
CA

Why it was recalled

The femoral impactors have a higher than normal risk of the arm breaking. During impaction the tip of the arm may fracture and separate from the instrument. This may cause a minor delay in surgery. If not retrieved, it may cause an inflammatory response that may require revision surgery.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

REF 800-05-035, ENZTEC FEMORAL IMPACTOR - DJO, (01) 09421028110624 (10) 999999, To be used only with DJO EMPOWR 3D KNEE