Medical device recalls Moderate risk

ErgoSafe Products, LLC (DBA) Prism Medical recalls A-series lift - A625 Manual Part No. 324117. Prism Medical.

Recall date
January 30, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1406-2017
FDA classification
Class II
Brand / firm
ErgoSafe Products, LLC (DBA) Prism Medical
Sold / distributed
Product was distributed to MT and NV; one VA consignee.

Why it was recalled

During internal cycle test at max load (625 lbs.) 2 units had straps break prematurely.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

A-series lift - A625 Manual Part No. 324117. Prism Medical.