Medical device recalls High risk

Ethicon Endo-Surgery Inc recalls SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An integrated physiological monitoring and drug deliv…

Recall date
May 13, 2015
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2374-2015
FDA classification
Class I
Brand / firm
Ethicon Endo-Surgery Inc
Sold / distributed
Nationwide Distribution including OH, RI, TX & WA.

Why it was recalled

Ethicon has found that disinfecting methods not specified in the Instructions For Use (IFU) have led to the degradation of the plastic material of the SEDASYS System Control Unit.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

SEDASYS Computer-Assisted Personalized Sedation System, SEDPRU01. An integrated physiological monitoring and drug delivery system intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures. Intended to provide physician-led teams with a means of administering 1% propofol (10 mg/mL) injectable emulsion to sedate patients > 18 (ASA I-II) undergoing esophagogastroduodenoscopy (EGD) and colonoscopy procedures.

Get recall alerts

Free email alert whenever Ethicon Endo-Surgery Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ethicon Endo-Surgery Inc