Medical device recalls Moderate risk

Ethicon, LLC recalls PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general s…

Recall date
December 20, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1096-2025
FDA classification
Class II
Brand / firm
Ethicon, LLC
Sold / distributed
Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, LA MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and the countries of AT, AU, BE, CA, CH, CN, CZ, DE, DK, FI, FR, GB, HU, IE…

Why it was recalled

Sutures may have an open seal on the primary packaging, due to a manufacturing issue on a specific machine, and if issue is undetected could lead to breach in sterility, could introduce pathogens to the patient and cause infection.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

PROLENE Polypropylene Suture, REF: EH7585H, 8706H, 8833H, 8963H, 8935H PROLENE Suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic, and neurosurgical procedures.

Get recall alerts

Free email alert whenever Ethicon, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Ethicon, LLC