FHC, Inc. recalls FHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulate electrophysiological…
- Recall date
- June 17, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2540-2020
- FDA classification
- Class II
- Brand / firm
- FHC, Inc.
- Sold / distributed
- US Nationwide distribution including in the states of CA, NC, TN, TX, WA.
Why it was recalled
An electrical sub-circuit was not properly connected. Its purpose is to eliminate the possibility of interference with proper performance of the Guideline 5 Main Processing Unit, if an ESD event (greater than 6kV) were to occur.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
FHC Guideline 4000 5.0 GL5 Main Processing Unit - Product Usage: intended to record and stimulate electrophysiological activity, as well as aid in the accurate placement of electrodes and other instruments. Catalogue Number: C0215 UDI: 00873263006740