Medical device recalls Moderate risk

Flexicare Medical Ltd. recalls BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Steril…

Recall date
November 27, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0804-2020
FDA classification
Class II
Brand / firm
Flexicare Medical Ltd.
Sold / distributed
US: Nationwide OUS: Australia, Canada, China, Czech Republic, Denmark, Finland, Ireland, Italy, Kuwait, Lebanon, Netherlands, Panama, Romania, Russia, South Korea, UAE, UK

Why it was recalled

The firm has received reports that the spring/washer/bearing components in the block of the laryngoscope blade become loose and separated from the device. If the blade is disengage over the patient's mount it is a possibility that any loose parts could enter the oral cavity which could cause a delay in the procedure while retrieving loose parts.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BriteBlade Pro Single-Use Fiber Optic Handle and Blade Miller 2, CE 0120, Do not Re-use, Rx Only, REF: 040-342U, Sterile EO, Qty 10, UDI: (01)15055788724053(17)230901(10)181002320

Get recall alerts

Free email alert whenever Flexicare Medical Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Flexicare Medical Ltd.