Flexicare Medical Ltd. recalls ProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U/ 7.0, 038-990…
- Recall date
- September 22, 2020
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0497-2021
- FDA classification
- Class I
- Brand / firm
- Flexicare Medical Ltd.
- Sold / distributed
- US nationwide distribution.
Why it was recalled
When the video stylet cuff is over-inflated there is potential for occlusion of the corrugated section of the ETT, and increased resistance to gas flow and difficulty in ventilating the patient.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
ProVu Single Use Video Stylet with ET Tube, Part/Size: 038-990-080U/ 8.0, 038-990-075U/ 7.5, 038-990-070U/ 7.0, 038-990-065U/ 6.5 *Note: product list updated per 12/23/2020 letter to customers
Get recall alerts
Free email alert whenever Flexicare Medical Ltd. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Flexicare Medical Ltd.