Medical device recalls Moderate risk

Flower Orthopedics Corporation recalls Flex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue R…

Recall date
November 2, 2021
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0448-2022
FDA classification
Class II
Brand / firm
Flower Orthopedics Corporation
Sold / distributed
US Nationwide distribution in the states of AZ, CA, CO, GA, TN, FL, IL, MA, MD, NC, TX.

Why it was recalled

Mating incompatibility between the Inserter Shaft and the Inserter Draw Rod causing the inserter draw rod to bind and not rotate freely as intended surgeon may be unable to complete surgery unless a second set is available

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Flex-Thread Inserter Shaft intended for Fibula intramedullary-intended to be used with the Inserter Draw Rod and Blue Ratchet Handle in order to facilitate the insertion of the Flex-Thread Nail System- Catalog Number: 8435-1