Medical device recalls Moderate risk

Flowonix Medical Inc recalls Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra I…

Recall date
March 21, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1712-2019
FDA classification
Class II
Brand / firm
Flowonix Medical Inc
Sold / distributed
US nationwide distribution.

Why it was recalled

The pump shuts down unexpectedly if Error 115, an alarm associated with a watchdog timer timeout, is triggered. There has been an increase of Error 115 occurrences due to a software defect caused by a Healthcare Provider using a Clinician Programmer, Software Version 2.00.29, to program a Bridge or Demand bolus while the pump is set to either a Periodic Flow or Multiple Rates flow mode.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Prometra Clinician Programmer (Touchscreen) Software version 2.00.29, Cat. No. 12828, used with Prometra and Prometra II Programmable Pumps. For intrathecal infusion.

Get recall alerts

Free email alert whenever Flowonix Medical Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Flowonix Medical Inc