Medical device recalls Moderate risk

Flowonix Medical Inc recalls Prometra Programmable 20mL Pump. Catalog No. 11827 The device is an implantable infusion pump. GTIN-14 00810335020082

Recall date
December 24, 2019
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1120-2020
FDA classification
Class II
Brand / firm
Flowonix Medical Inc
Sold / distributed
Nationwide domestic distribution. Foreign distribution to United Kingdom, Republic of Ireland, Poland, Greece, Italy, Germany, Netherlands, and Belgium.

Why it was recalled

A pump alarm function anomaly in the pump firmware code may result in the pump not sending the expected error code notifications to the Programmer, or not emit an audible alarm tone in certain "out of tolerance" conditions associated with the pump's valve system.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Prometra Programmable 20mL Pump. Catalog No. 11827 The device is an implantable infusion pump. GTIN-14 00810335020082