Medical device recalls Moderate risk

Fresenius Medical Care Renal Therapies Group, LLC recalls 2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840…

Recall date
December 14, 2017
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1278-2018
FDA classification
Class II
Brand / firm
Fresenius Medical Care Renal Therapies Group, LLC
Sold / distributed
US Distribution to the states of: AR, CA, CO, FL, GA, IL, MA, MD, MN, NC, NY, OH, PA, SC, TN, and TX.

Why it was recalled

When the recirculation ultrafiltration (UF) Goal is set to a value greater than 200ml in service mode and the user starts a treatment using the SLED (Sustained Low Efficiency Dialysis) program the display will show an invalid message and the UF pump will run at the recirculation UF rate which may be up to 4000ml/hour. The SLED program is required to limit the UF rate to a maximum of 1000 ml/hour.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

2008 K2 Hemodialysis Machine with software version 5.40, Models: (1) Hemodialysis SYS OLC/Diasafe PLS 190610, UDI 00840861100859 (+Serial Number), (2) Machine Short Cap, OLC.DP HP 190630, UDI 00840861100866 (+Serial Number) ndicated for acute and chronic dialysis therapy.

Get recall alerts

Free email alert whenever Fresenius Medical Care Renal Therapies Group, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Fresenius Medical Care Renal Therapies Group, LLC